The synthesis cycle has been written for position 5 on the Expedite™. If any other positions are used, third line in the coupling step has to be changed.
/* ------------------------------------------------------------------------------------ */
/* Function Mode Amount Time(sec) Description */
/* /Arg1 /Arg2 */
/* ------------------------------------------------------------------------------------ */
$Deblocking
12 /*Wsh A */PULSE 15 0 "Pre cycle wash"
144 /*Index Fract. Coll.*/NA 1 0 "Event out ON"
0 /*Default */WAIT 0 1.5 "Wait"
16 /*Dblk */PULSE 20 0 "Dblk to column"
141/*Trityl Mon. On/Off*/ NA 1 1 "START data collection"
16 /*Dblk */PULSE 20 0 "Deblock"
16 /*Dblk */ PULSE 5 10 "Deblock"
16 /*Dblk */ PULSE 10 0 "Deblock"
16 /*Dblk */ PULSE 5 10 "Deblock"
16 /*Dblk */ PULSE 10 0 "Deblock"
16 /*Dblk */ PULSE 5 10 "Deblock"
16 /*Dblk */ PULSE 10 0 "Deblock"
38 /*Diverted Wsh A */ PULSE 20 20 "Deblock"
38 /*Diverted Wsh A */ PULSE 40 0 "Flush system with Wsh A"
141/*Trityl Mon. On/Off */ NA 0 1 "STOP data collection"
144 /*Index Fract. Coll. */ NA 2 0 "Event out OFF"
$Coupling
1 /*Wsh */ PULSE 10 0 "Flush system with Wsh"
2 /*Act */ PULSE 5 0 "Flush system with Act"
22 /*5 + Act */ PULSE 5 0 "Monomer + Act to column"
2 /*Act */ PULSE 5 0 "Chase with Act"
1 /*Wsh */ PULSE 12 0 "Chase with Wsh"
1 /*Wsh */ PULSE 10 250 "Slow pulse to couple"
1 /*Wsh */ PULSE 4 0 "Flush with Wsh"
$Capping
13 /*Caps */ PULSE 12 0 "Caps to column"
12 /*Wsh A */ PULSE 10 0 "Chase with Wsh A"
12 /*Wsh A */ PULSE 10 45 "Slow pulse to cap"
$Oxidizing
15 /*Ox */ PULSE 15 0 "Ox to column"
12 /*Wsh A */ PULSE 12 0 "Chase with Wsh A"
12 /*Wsh A */ PULSE 10 45 "Slow pulse to Ox"
$Capping
13 /*Caps */ PULSE 10 0 "Caps to column"
12 /*Wsh A */ PULSE 10 0 "Chase with Wsh A"
12 /*Wsh A */ PULSE 10 10 "Slow pulse to Cap"
12 /*Wsh A */ PULSE 20 0 "End of cycle wash"
Bottle/position codes:
18 /*A + Act */
19 /*C + Act */
20 /*G + Act */
21 /*T + Act */
22 /*5 + Act */
23 /*6 + Act */
24 /*7 + Act */
25 /*8 + Act */
26 /*9 + Act */
The synthesis cycle has been written for position A on the Expedite™. If any other positions are used, third line in the coupling step has to be changed.
MOSS DNA-cycle (0.2μmole scale)
/* ---------------------------------------------------------------------------------- */
/* Function Mode Amount Time(sec) Description */
/* /Arg1 /Arg2 */
/* ---------------------------------------------------------------------------------- */
$Deblocking
12 /*Wsh A */ PULSE 15 0 "Pre cycle wash"
144/*Index Fract. Coll.*/ NA 1 0 "Event out ON"
0 /*Default */ WAIT 0 1.5 "Wait"
16 /*Dblk */ PULSE 20 0 "Dblk to column"
141/*Trityl Mon. On/Off*/ NA 1 1 "START data collection"
16 /*Dblk */ PULSE 20 0 "Deblock"
16 /*Dblk */ PULSE 5 10 "Deblock"
16 /*Dblk */ PULSE 10 0 "Deblock"
16 /*Dblk */ PULSE 5 10 "Deblock"
16 /*Dblk */ PULSE 10 0 "Deblock"
16 /*Dblk */ PULSE 5 10 "Deblock"
16 /*Dblk */ PULSE 10 0 "Deblock"
38 /*Diverted Wsh A */ PULSE 20 20 "Deblock"
38 /*Diverted Wsh A */ PULSE 40 0 "Flush system with Wsh A"
141/*Trityl Mon. On/Off*/ NA 0 1 "STOP data collection"
144/*Index Fract. Coll. */ NA 2 0 "Event out OFF"
$Coupling
1 /*Wsh */ PULSE 10 0 "Flush system with Wsh"
2 /*Act */ PULSE 5 0 "Flush system with Act"
18 /*A + Act */ PULSE 5 0 "Monomer + Act to column"
2 /*Act */ PULSE 5 0 "Chase with Act"
1 /*Wsh */ PULSE 12 0 "Chase with Wsh"
1 /*Wsh */ PULSE 10 120 "Slow pulse to couple"
1 /*Wsh */ PULSE 4 0 "Flush with Wsh"
$Capping
13 /*Caps */ PULSE 12 0 "Caps to column"
12 /*Wsh A */ PULSE 10 0 "Chase with Wsh A"
12 /*Wsh A */ PULSE 10 45 "Slow pulse to cap"
$Oxidizing
15 /*Ox */ PULSE 15 0 "Ox to column"
12 /*Wsh A */ PULSE 12 0 "Chase with Wsh A"
12 /*Wsh A */ PULSE 10 30 "Slow pulse to Ox"
$Capping
13 /*Caps */ PULSE 10 0 "Caps to column"
12 /*Wsh A */ PULSE 10 0 "Chase with Wsh A"
12 /*Wsh A */ PULSE 10 10 "Slow pulse to Cap"
12 /*Wsh A */ PULSE 20 0 "End of cycle wash"
Bottle/position codes:
18 /*A + Act */
19 /*C + Act */
20 /*G + Act */
21 /*T + Act */
22 /*5 + Act */
23 /*6 + Act */
24 /*7 + Act */
25 /*8 + Act */
26 /*9 + Act */
Trademarks and patents
Exiqon™ is a registered trademark of Exiqon A/S, Vedbaek, Denmark. Locked Nucleic Acid (LNA™) is covered by patents/patents applications and corresponding worldwide applications owned by Exiqon A/S and Prof. Imanishi. Expedite™ is a trademark of Applied Biosystems.
Latest revision May 30, 2005